Measurement Capability
| Parameter / Test | Range / Accuracy / Specifications |
|---|---|
| Triple Point of Mercury ( -38.8344 Degree C) | ± 2.83 mK |
| Triple Point of Water ( 0.01 Degree C) | ± 2.00 mK |
| Melting Point of Gallium ( 29.7646 Degree C) | ± 2.10 mK |
| Freezing Point of Tin ( 231.928 Degree C) | ± 2.50 mK |
| Freezing Point of Zinc ( 419.527 Degree C) | ± 5.00 mK |
| Freezing Point of Aluminum ( 660.323 Degree C) | ± 7.20 mK |
List of Major Facility
| Facility / Equipment / Tools | Make and Model |
|---|---|
| Triple Point of Mercury Cell | Isotech / 17724 |
| Triple Point of Water Cell | Hart Sc. / 5901 |
| Melting Point of Gallium Cell | Hart Sc. / 5943 |
| Freezing Point of Tin Cell | Isotech / 17669 |
| Freezing Point of Zinc Cell | Isotech / 17671 |
| Freezing Point of Aluminum Cell | Isotech / 17672 |
| 4 -Terminal Standard Resistor | Guildline / 7334 |
| Variable Temperature Air Bath | Guildline / 5032 |
| Precision DCC Temperature - Resistance Bridge | Guildline / 6622B |
| Standard Platinum Resistance Thermometer, 25.5 Ω | Hart Sc. / 5681 |
Major Products Tested / Project Done
| Major Products Tested / Projects Done |
|---|
Quartz glass sheathed, long stem SPRT / PRT sensors * From - 38.8344 to 660.323 Degree C Metal sheathed, long stem SPRT / PRT / RTD sensors * From - 38.8344 to 419.527 Degree C * Diameter of sensor should be ≤ 8mm and it's length should be ≥ 380 mm. |